What is a Design History File?
The Design History File (DHF) is a comprehensive compilation of documentation that provides a detailed record of the design and development of a medical device. It is required by regulatory bodies such as the U.S. Food and Drug Administration (FDA) under the Quality System Regulation (21 CFR Part 820) and serves as a means to demonstrate compliance with design control requirements. The DHF should include:
- Design and development plans that outline the activities involved in the design process.
- Design inputs, including the physical and performance requirements of the device.
- Design outputs, which are the results of the design process and include specifications, drawings, and algorithms.
- Design reviews, including dates, participants, and outcomes of formal reviews during the design process.
- Design verification and validation documentation showing that the design outputs meet the design inputs and that the device meets user needs and intended uses.
- Records of design changes and how they were reviewed, verified, validated, and approved.
- Any additional documentation generated throughout the design and development process.
The DHF is a living document that must be maintained and updated throughout the life of the medical device to reflect any design changes. It is a critical component of a medical device’s quality system and is subject to review during FDA inspections and audits by Notified Bodies in the EU.